a p p l i c at i o n n o t e

                                                              Gas Chromatography/
                                                              Mass Spectrometry

                                                              Authors
                                                              Yury Kaplan
                                                              Ruben Garnica
                                                              PerkinElmer, Inc.
                                                              Shelton, CT 06484 USA




Improved Sensitivity                          Introduction
                                                 U.S. Environmental Protection
and Dynamic Range                                Agency (EPA) Method 8270D – Semi-
                                                 Volatile Organic Compounds by Gas
Using the Clarus SQ 8                            Chromatography/Mass Spectrometry
                                                 (GC/MS) – is a common and wide ranging
GC/MS System for                                 method employed in nearly all commercial
                                                 environmental laboratories. The analysis
EPA Method 8270D                                 focuses on the detection of trace level
                                                 semi-volatile organic compounds in
Semi-Volatile Organic                            extracts from solid waste matrices, soils,
                                                 air sampling media and water samples.
Compound Analysis                                The method lists over 200 compounds
                                                 however a majority of laboratories target
                                                 between 60 and 90 for most analyses.
                                                 The study presented here demonstrates
      the PerkinElmer® Clarus® SQ 8 GC/MS, not only meets the method requirements
      but provides users flexibility to satisfy their individual productivity demands.
      An extended calibration range is presented as are the advantages of the Clarifi™
      detector.
Experimental                                                     Calibration and performance standards were prepared
    The PerkinElmer Clarus SQ 8T GC/MS system operating in           from commercially available stock standards and diluted in
    electron ionization mode was used to perform these experi-       Class-A volumetric flasks. Calibration standards were made
    ments. Experimental conditions are presented in Table 1          from a 76 component 8270 mega mix, a 15 component
    and represent standard conditions for EPA Method 8270D.          acids fortification mix and a 6 component semi-volatile
    Minimum inlet reactivity was achieved using a deactivated        internal standard mix. The acids were fortified with the
    quartz liner (4-mm) without wool. As prescribed by the           fortification mix to double the concentration relative to
    method, a pressure pulse injection scheme was employed           the other components. Stock standards were diluted to
    and is described in Table 2. Data collection was achieved        an intermediate concentration from which initial calibra-
    using the TurboMass™ v6 GC/MS software with environ-             tion standards were prepared. The calibration levels used
    mental reporting. The TurboMass software comes with a            in this study are presented in Table 3. The internal standard
    built-in UltraTune™ (Standard – DFTPP/BFB) tuning procedure,     mix was added to all samples such that a resulting concen-
    which was used to establish the base detector setting.           tration of 40 μg/mL was achieved. The tuning standard,
                                                                     Decafluorotriphenylphosphine (DFTPP), was prepared at
                                                                     50 μg/mL. Standards, stock standards, and associated
     Table 1. GC and MS experimental conditions.
                                                                     QC/QA samples were stored in a manner consistent with
     Analytical Column:    Elite-5MS (30 m x 0.25 mm id x 0.25 μm)   the guidelines set out in the method.
     Injector Temperature: 300 °C
     Carrier Gas:          Helium @ 1.0 mL/min                        Table 3. Calibration points employed in this study.
     Oven Program:         Temperature Hold Time          Rate                               “Acids”             “Bases/Neutrals”
                               37 °C         0.5 min    18 °C/min     Calibration         Concentration           Concentration
                                                                       Level                (μg/mL)                  (μg/mL)
                               265 °C         0 min     6 °C/min
                                                                           1                    2                           1
                               287 °C         0 min     25°C/min
                                                                           2                    4                           2
                               320 °C        1.85 min      End
                                                                           3                    8                           4
     GC Transfer Line
     Temperature:          280 °C                                          4                    12                          6

     MS Ion Source                                                         5                    20                      10
     Temperature:          280 °C                                          6                    30                      15
     MS Function Type:     Full Scan                                       7                    40                      20
     MS Solvent Delay:     0 – 2.30 min                                    8                    80                      40
     MS Scan Range:        m/z 35 – 500                                    9                   160                      80
     MS Scan Time:         0.10 sec                                        10                  200                      100
     MS Interscan Delay:   0.10 sec                                        11                  300                      150



     Table 2. Pressure plus flow settings.                           Results and Discussion
     Time (min)            Split Flow (mL/min)                       An initial consideration when performing any method is to
                                                                     determine the operating range with respect to on-column
     -0.50                             4
                                                                     concentration. The Clarus SQ 8 GC/MS allows the user to
     -0.25                             0                             control the linear range by controlling the Clarifi detector
     0.75                              1                             voltage setting. The built-in UltraTune (Standard – DFTPP/
     1.00                             50                             BFB) tune procedure adjusts the mass spectrometer providing
                                                                     excellent spectral fidelity and sets the detector voltage such
     5.00                             20
                                                                     that low sensitivity work can be performed. This standard




2
setting can be used directly and is recommended for most                                                                                                                                                                             Scan EI+
                                                                                                                                                    198                                                                                7.22e8
applications. Additionally, the detector voltage can be                                        100


increased for added sensitivity when needed, or, as is the
case with this application note, the detector voltage can be
tuned to provide both sensitivity and a wide dynamic range                                                                                                                                                         442




                                                                                                                77
                                                                                                %
                                                                                                                                127                                   255
                                                                                                               69


                                                                                                      51
                                                                                                                          110
                                                                                                                                                                            275
                                                                                                                                                        206
                                                                                                                                  129
                                                                                                                      107                         186                                                                443
                                                                                                                                                                                                               441
                                                                                                                                                              224
                                                                                                     50                                                             244 256
                                                                                                                            117
                                                                                                     39             78 93                   167                                   296                        423
                                                                                                          57                          148                                                323 334 365                 444
                                                                                                 0                                                                                                                                          m/z
                                                                                                     43              93            143             193              243           293       343        393         443         493    543



                                                                                               Figure 2. Mass spectrum of DFTPP as analyzed by the tune evaluation shown
                                                                                               in Figure 1.


                                                                                               suitable for the analysis at hand. In this study the Clarifi
                                                                                               detector voltage was optimized such that both the low level
                                                                                               samples were measurable while the high level samples were
                                                                                               not saturating the detector. Specifically, the detector was
                                                                                               lowered 260V to the operational 1400V. The added sensitivity
                                                                                               of the Clarifi detector enhances a wider range of analyses
                                                                                               so that even at low voltage settings the low concentration
Figure 1. EPA 8270D tune evaluation as performed by the built-in software.
Passing result shown.                                                                          standards and samples are easily detected.


                                                                                                                                                                                                                              Scan EI+
                                                                                                                                            63                                                                                     TIC
  100

                                                                      43




                                                             32 36
                                      23
                                                                                                     57 58                               62
                                                                                  54
                                                                                53
                                               25                      46                                                       60
                                     20             27 28                                                                                                                                                          71
                                                                                52
                                                       29        38
                   2                                                                                                                                                          67                                         72
                                                                           47
                        12                24                                              56                              59
    %                                                                      48 50
                       8 13                                                                                                                                                             68 69
                                18                                41           51
                       10 15                                                                                                                                        65
                                                                                                                                                                          66
                                                            31

                           16                                                        55

                              17                        30




        1

     0                                                                                                                                                                                                                           Time
     3.00              5.00                    7.00               9.00                 11.00                          13.00                                   15.00                             17.00                         19.00


Figure 3. TIC of 40 μg/mL calibration sample. Retention Times are listed in Table 4 (Page 4).




                                                                                                                                                                                                                                                  3
67
                                                                                                   Scan EI+
                                                                                                        252
                                                                                                              In addition to providing a solid foundation from which to
     100
                                                                                                              perform analyses the UltraTune (Standard – DFTPP/BFB)
                                                                                                              function also produces a satisfactory tune, which meets the
                                                                    68
                                                                                                              tune evaluation requirements set out in Method 8270D.
                                      66
                                                                                                              Figure 1 demonstrates the passing DFTPP tune evaluation
                                                                                                              sample while Figure 2 shows the mass spectrum utilized in
      %
                                                                                                              the tune evaluation test.

                                                                                                              The GC conditions were optimized to provide analytical
                                                                                                              separation and Figure 3 demonstrates a mid-level (40 μg/mL)
                                                                                                              total ion chromatogram (TIC) from this analysis. All of the
                                                                                                              expected analyte separation is achieved and a listing of
                                                                                                              retention times (RT) measured using the instrument parameters
      0                                                                                               Time
      15.00   15.20   15.40   15.60         15.80   16.00   16.20        16.40   16.60     16.80   17.00
                                                                                                              as described in Table 1 is presented. The Elite-5MS column
    Figure 4. Zoom of EIC at m/z = 252 showing separation of benzo[b]                                         provided excellent separation and Figure 4, for example,
    fluoranthene (66) and benzo[k]fluoranthene (67). Benzo[a]pyrene (68)                                      demonstrates the separation of the benzo[b]fluoranthene
    also shown.                                                                                               and benzo[k]fluoranthene.



      Table 4. Calibration data table with retention times listed in Figure 3 (Page 3).
      #           Name                                                           RT (min)                     Average RRF         8270D Criteria RRF          % RSD
      1           Pyridine                                                               3.09                     0.2                                           8.98
      2           Aniline                                                                4.44                     1.2                                          13.16
      3           Phenol                                                                 4.45                     1.7                     0.8                   9.52
      4           N-Nitrosodimethylamine                                                 4.45                     0.1                                           10.3
      5           Bis (2-chloroethyl) ether                                              4.49                     2.1                     0.7                  12.66
      6           1,3-Dichlorobenzene                                                    4.66                     1.1                                          15.56
      7           1,4-Dichlorobenzene - D4 (ISTD)                                        4.73                                                                   8.83
      8           1,4-Dichlorobenzene                                                    4.75                     1.0                                           9.73
      9           1,2-Dichlorobenzene                                                    4.91                     1.3                                          11.27
      10          Benzyl Alcohol                                                         4.94                     1.0                                          11.86
      11          Bis (2-chloroisopropyl) ether                                          5.05                     1.7                                           9.34
      12          2-Methylphenol                                                         5.07                     0.6                     0.7                  10.66
      13          3-Methylphenol                                                         5.26                     0.7                                          11.71
      14          4-Methylphenol                                                         5.26                     0.9                     0.6                   8.59
      15          Hexachloroethane                                                       5.28                     0.5                     0.3                  10.62
      16          Nitrobenzene                                                           5.41                     0.4                     0.2                   9.91
      17          Isophorone                                                             5.70                     0.7                     0.4                  11.31
      18          2-Nitrophenol                                                          5.78                     0.2                                          11.86
      19          Bis(2-chloroethoxy)methane                                             5.96                     0.5                     0.3                  15.45
      20          2,4-Dichlorophenol                                                     6.08                     0.3                     0.2                  16.69
      21          1,2,4-Trichlorobenzene                                                 6.14                     0.4                                          14.19
      22          Naphthalene - D8 (ISTD)                                                6.21                                                                   7.41
      23          Naphthalene                                                            6.24                     1.1                     0.7                   9.63
      24          Hexachlorobutadiene                                                    6.39                     0.2                    0.01                  10.38
      25          4-chloro-3-methyl-phenol                                               7.01                     0.3                     0.2                  16.86
      26          2-Methylnaphthalene                                                    7.09                     0.6                     0.4                  14.73
      27          1-Methylnaphthalene                                                    7.22                     0.6                                          13.96
      28          Hexachlorocyclopentadiene                                              7.30                     0.2                    0.05                   9.14
      29          2-Chloronaphthalene                                                    7.72                     1.4                     0.8                  12.09

4
Table 4 continued
#       Name                           RT (min)   Average RRF   8270D Criteria RRF   % RSD
30      2-Nitroaniline                   7.93         0.4              0.01           9.54
31      Dimethyl phthalate               8.19         1.4              0.01          10.67
32      2,6-Dinitrotoluene               8.25         0.3              0.2           13.95
33      Acenaphthylene                   8.25         2.0              0.9            8.27
34      Acenaphthene - D10 (ISTD)        8.45                                         7.56
35      Acenaphthene                     8.49         1.5              0.9            9.85
36      3-Nitroaniline                   8.49         0.2              0.01           14.4
37      2,4-Dinitrophenol                8.60         0.1              0.01           8.17
38      Dibenzofuran                     8.71         1.7              0.8           11.83
39      4-Nitrophenol                    8.79         0.1              0.01          10.14
40      2,3,4,6-Tetrachlorophenol        8.85         0.2                            12.71
41      2,3,5,6-Tetrachlorophenol        8.91         0.2                            10.17
42      Diethylphthalate                 9.09         1.1              0.01           9.93
43      Fluorene                         9.15         1.3              0.9           11.93
44      4-Chlorodiphenylether            9.17         0.5              0.4           13.53
45      4,6-Dinitro-2-methylphenol       9.29         0.1              0.01          16.04
46      4-Nitroaniline                   9.31         0.1              0.01          12.33
47      N-Nitrosodiphenylamine           9.36         0.5              0.01           11.4
48      4-Bromophenyl phenyl ether       9.80         0.2              0.1            9.57
49      Azobenzene                       9.81         0.1                            10.13
50      Hexachlorobenzene                9.88         0.2              0.1           10.17
51      Pentachlorophenol               10.18         0.1              0.05           8.01
52      Phenathrene - D10 (ISTD)        10.38                                         3.48
53      Phenanthrene                    10.42         1.1              0.7            13.5
54      Anthracene                      10.48         1.1              0.7           15.16
55      Carbazole                       10.72         0.4              0.01           8.39
56      Di-n-butyl phthalate            11.20         0.8              0.01          11.42
57      Fluoroanthene                   11.97         1.0              0.6            8.52
58      Pyrene                          12.26         1.2              0.6            9.62
59      Butyl benzyl phthalate          13.16         0.4              0.01          12.35
60      Bis (2-ethylhexyl) adipate      13.26         0.5                            14.38
61      Benzo[a]anthracene              13.89         1.0              0.8            11.8
62      Chrysene - D12 (ISTD)           13.92                                         6.34
63      Chrysene                        13.96         1.0              0.7           16.57
64      Bis (2-ethylhexyl) phthalate    13.98         0.6              0.01          17.47
65      Di-n-Octyl phthalate            15.04         1.0                            11.13
66      Benzo[b]fluoranthene            15.68         1.2              0.7           12.55
67      Benzo[k]fluoranthene            15.74         0.9              0.7           13.52
68      Benzo[a]pyrene                  16.29         1.0              0.7            9.46
69      Perylene - D12 (ISTD)           16.42                                         4.24
70      Indeno(1,2,3-c,d)pyrene         18.00         1.2              0.5            8.72
71      Dibenz[a,h]anthracene           18.12         1.0              0.4            9.25
72      Benzo[g,h,i] perylene           18.42         1.0              0.5            7.98



                                                                                             5
The results of the initial calibration curve for EPA Method
8270D are presented in Table 4 and include retention time,
average relative response factor (Avg. RRF), method criteria,
and percent relative standard deviation (% RSD). Satisfactory
results are obtained for all for the compounds listed.


Conclusion
In this application note the Clarus SQ 8 GC/MS system is
shown to provide wide concentration range applicability to
U.S. EPA Method 8270D, allowing the analysts flexibility
in performing experimentation to meet their needs. Satisfactory
analytical results were demonstrated over a concentration
range from 1.0 – 150 μg/mL for a majority of analyte
compounds using the full scan approach providing library
searchable spectra at all concentration levels. A number
of technological advances make the Clarus SQ 8 GC/MS
the ideal systems for laboratories wishing to perform high
throughput and sensitivity analyses with an ease of operation
that is currently unmatched.




PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com



For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.

010016_01

Mais conteúdo relacionado

Mais procurados

Guia de TGA (análisis termogravimétrico)
Guia de TGA (análisis termogravimétrico)Guia de TGA (análisis termogravimétrico)
Guia de TGA (análisis termogravimétrico)Alan Viveros
 
Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...
Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...
Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...PerkinElmer, Inc.
 
Calorimetry study of a mAb that precipitates upon thermal denaturation
Calorimetry study of a mAb that precipitates upon thermal denaturationCalorimetry study of a mAb that precipitates upon thermal denaturation
Calorimetry study of a mAb that precipitates upon thermal denaturationKBI Biopharma
 
Edwards - qNMR Prresentation - Aloe Summit - Nov 2012
Edwards - qNMR Prresentation - Aloe Summit - Nov 2012Edwards - qNMR Prresentation - Aloe Summit - Nov 2012
Edwards - qNMR Prresentation - Aloe Summit - Nov 2012John Edwards
 
Differential Scanning Calorimeter and Recent Advances
Differential Scanning Calorimeter and Recent Advances Differential Scanning Calorimeter and Recent Advances
Differential Scanning Calorimeter and Recent Advances Joachim Emeka Arikibe
 
Mdo2014 takeda
Mdo2014 takedaMdo2014 takeda
Mdo2014 takedaEd Tessier
 
GAS CHROMATOGRAPHY-MASS SPECTOMETRY
GAS CHROMATOGRAPHY-MASS SPECTOMETRYGAS CHROMATOGRAPHY-MASS SPECTOMETRY
GAS CHROMATOGRAPHY-MASS SPECTOMETRYAmeemahkhan1
 
Gas chromatography in environmental analysis driscoll 2004 l
Gas chromatography in environmental analysis  driscoll  2004 lGas chromatography in environmental analysis  driscoll  2004 l
Gas chromatography in environmental analysis driscoll 2004 lDr. Jack Driscoll
 
Thermal analysis characterization of polymers and plastics acs webinar
Thermal analysis characterization of polymers and plastics acs webinarThermal analysis characterization of polymers and plastics acs webinar
Thermal analysis characterization of polymers and plastics acs webinarKevin Menard, Ph.D. MBA
 
Simplified Cannabis Terpene Profiling by GCMS
Simplified Cannabis Terpene Profiling by GCMSSimplified Cannabis Terpene Profiling by GCMS
Simplified Cannabis Terpene Profiling by GCMSShinji Nakamura
 

Mais procurados (20)

New Innovations in Ultra High Performance Liquid Chromatography and Liquid Ch...
New Innovations in Ultra High Performance Liquid Chromatography and Liquid Ch...New Innovations in Ultra High Performance Liquid Chromatography and Liquid Ch...
New Innovations in Ultra High Performance Liquid Chromatography and Liquid Ch...
 
Guia de TGA (análisis termogravimétrico)
Guia de TGA (análisis termogravimétrico)Guia de TGA (análisis termogravimétrico)
Guia de TGA (análisis termogravimétrico)
 
Dsc application 2
Dsc application 2Dsc application 2
Dsc application 2
 
Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...
Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...
Application Note:Determination of Impurities in Semiconductor-Grade TMAH with...
 
Calorimetry study of a mAb that precipitates upon thermal denaturation
Calorimetry study of a mAb that precipitates upon thermal denaturationCalorimetry study of a mAb that precipitates upon thermal denaturation
Calorimetry study of a mAb that precipitates upon thermal denaturation
 
RADTEC98_final
RADTEC98_finalRADTEC98_final
RADTEC98_final
 
Edwards - qNMR Prresentation - Aloe Summit - Nov 2012
Edwards - qNMR Prresentation - Aloe Summit - Nov 2012Edwards - qNMR Prresentation - Aloe Summit - Nov 2012
Edwards - qNMR Prresentation - Aloe Summit - Nov 2012
 
Chromatography: Trends and Developments in MAb Screening and Characterization
Chromatography: Trends and Developments in MAb Screening and CharacterizationChromatography: Trends and Developments in MAb Screening and Characterization
Chromatography: Trends and Developments in MAb Screening and Characterization
 
Cat 1
Cat 1Cat 1
Cat 1
 
Flexible Automated Sample Prep Workflows
Flexible Automated Sample Prep WorkflowsFlexible Automated Sample Prep Workflows
Flexible Automated Sample Prep Workflows
 
Differential Scanning Calorimeter and Recent Advances
Differential Scanning Calorimeter and Recent Advances Differential Scanning Calorimeter and Recent Advances
Differential Scanning Calorimeter and Recent Advances
 
Mdo2014 takeda
Mdo2014 takedaMdo2014 takeda
Mdo2014 takeda
 
Water Analysis using LAMBDA
Water Analysis using LAMBDAWater Analysis using LAMBDA
Water Analysis using LAMBDA
 
Practical concerns in ega
Practical concerns in egaPractical concerns in ega
Practical concerns in ega
 
Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water
 
GAS CHROMATOGRAPHY-MASS SPECTOMETRY
GAS CHROMATOGRAPHY-MASS SPECTOMETRYGAS CHROMATOGRAPHY-MASS SPECTOMETRY
GAS CHROMATOGRAPHY-MASS SPECTOMETRY
 
Thermal analysis
Thermal analysisThermal analysis
Thermal analysis
 
Gas chromatography in environmental analysis driscoll 2004 l
Gas chromatography in environmental analysis  driscoll  2004 lGas chromatography in environmental analysis  driscoll  2004 l
Gas chromatography in environmental analysis driscoll 2004 l
 
Thermal analysis characterization of polymers and plastics acs webinar
Thermal analysis characterization of polymers and plastics acs webinarThermal analysis characterization of polymers and plastics acs webinar
Thermal analysis characterization of polymers and plastics acs webinar
 
Simplified Cannabis Terpene Profiling by GCMS
Simplified Cannabis Terpene Profiling by GCMSSimplified Cannabis Terpene Profiling by GCMS
Simplified Cannabis Terpene Profiling by GCMS
 

Semelhante a Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Method 8270D Semi-Volatile Organic Compound Analysis

Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...PerkinElmer, Inc.
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...PerkinElmer, Inc.
 
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer, Inc.
 
Cwea Me Hg Presentation Er
Cwea Me Hg Presentation ErCwea Me Hg Presentation Er
Cwea Me Hg Presentation Ernoel_enoki
 
SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014Kenneth Wong
 
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...KBI Biopharma
 
GHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIEGHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIEQian Xie
 
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Shimadzu Scientific Instruments
 
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i..."Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...Oscar Cabrices PhD
 
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...PerkinElmer, Inc.
 
sawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptxsawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptxEhabAdel29
 
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...PerkinElmer, Inc.
 
advance phytochemical analysis.pptx
advance phytochemical analysis.pptxadvance phytochemical analysis.pptx
advance phytochemical analysis.pptxPratikKapse8
 
Development and validation of new rp hplc method for analysis of capecitabine...
Development and validation of new rp hplc method for analysis of capecitabine...Development and validation of new rp hplc method for analysis of capecitabine...
Development and validation of new rp hplc method for analysis of capecitabine...IJSIT Editor
 
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...PerkinElmer, Inc.
 
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...PerkinElmer, Inc.
 
NETZSCH Nevio Instrument Series for THERMAL ANALYSIS
NETZSCH Nevio Instrument Series for  THERMAL ANALYSISNETZSCH Nevio Instrument Series for  THERMAL ANALYSIS
NETZSCH Nevio Instrument Series for THERMAL ANALYSISValentyn Mohylyuk
 

Semelhante a Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Method 8270D Semi-Volatile Organic Compound Analysis (20)

Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
 
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
 
Cwea Me Hg Presentation Er
Cwea Me Hg Presentation ErCwea Me Hg Presentation Er
Cwea Me Hg Presentation Er
 
GC-MS .pptx
GC-MS .pptxGC-MS .pptx
GC-MS .pptx
 
SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014
 
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
 
Increasing the Throughput of UHPLC
Increasing the Throughput of UHPLCIncreasing the Throughput of UHPLC
Increasing the Throughput of UHPLC
 
GHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIEGHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIE
 
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
 
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i..."Productivity and Simplicity" -  Streamlining Cumbersome Sample Preparation i...
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
 
EMSDVici
EMSDViciEMSDVici
EMSDVici
 
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
Determination of Impurities in Organic Solvents used in the Semiconductor Ind...
 
sawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptxsawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptx
 
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
 
advance phytochemical analysis.pptx
advance phytochemical analysis.pptxadvance phytochemical analysis.pptx
advance phytochemical analysis.pptx
 
Development and validation of new rp hplc method for analysis of capecitabine...
Development and validation of new rp hplc method for analysis of capecitabine...Development and validation of new rp hplc method for analysis of capecitabine...
Development and validation of new rp hplc method for analysis of capecitabine...
 
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
PerkinElmer: A Novel High Sensitivity Mass Spectrometric Detector for Gas Chr...
 
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
 
NETZSCH Nevio Instrument Series for THERMAL ANALYSIS
NETZSCH Nevio Instrument Series for  THERMAL ANALYSISNETZSCH Nevio Instrument Series for  THERMAL ANALYSIS
NETZSCH Nevio Instrument Series for THERMAL ANALYSIS
 

Mais de PerkinElmer, Inc.

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestrePerkinElmer, Inc.
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoPerkinElmer, Inc.
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!PerkinElmer, Inc.
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadePerkinElmer, Inc.
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no BrasilPerkinElmer, Inc.
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSPerkinElmer, Inc.
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorPerkinElmer, Inc.
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada PerkinElmer, Inc.
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESPerkinElmer, Inc.
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...PerkinElmer, Inc.
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...PerkinElmer, Inc.
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingPerkinElmer, Inc.
 

Mais de PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 

Último

Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24
Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24
Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24Mark Goldstein
 
Scale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL RouterScale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL RouterMydbops
 
Decarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a realityDecarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a realityIES VE
 
Moving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdfMoving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdfLoriGlavin3
 
Time Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directionsTime Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directionsNathaniel Shimoni
 
Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024TopCSSGallery
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
The State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptxThe State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptxLoriGlavin3
 
A Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptxA Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptxLoriGlavin3
 
[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality Assurance[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality AssuranceInflectra
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Farhan Tariq
 
MuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotes
MuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotesMuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotes
MuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotesManik S Magar
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024BookNet Canada
 
Zeshan Sattar- Assessing the skill requirements and industry expectations for...
Zeshan Sattar- Assessing the skill requirements and industry expectations for...Zeshan Sattar- Assessing the skill requirements and industry expectations for...
Zeshan Sattar- Assessing the skill requirements and industry expectations for...itnewsafrica
 
Glenn Lazarus- Why Your Observability Strategy Needs Security Observability
Glenn Lazarus- Why Your Observability Strategy Needs Security ObservabilityGlenn Lazarus- Why Your Observability Strategy Needs Security Observability
Glenn Lazarus- Why Your Observability Strategy Needs Security Observabilityitnewsafrica
 
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...Wes McKinney
 
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS:  6 Ways to Automate Your Data IntegrationBridging Between CAD & GIS:  6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integrationmarketing932765
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesKari Kakkonen
 
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxThe Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxLoriGlavin3
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsPixlogix Infotech
 

Último (20)

Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24
Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24
Arizona Broadband Policy Past, Present, and Future Presentation 3/25/24
 
Scale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL RouterScale your database traffic with Read & Write split using MySQL Router
Scale your database traffic with Read & Write split using MySQL Router
 
Decarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a realityDecarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a reality
 
Moving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdfMoving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdf
 
Time Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directionsTime Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directions
 
Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
The State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptxThe State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptx
 
A Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptxA Deep Dive on Passkeys: FIDO Paris Seminar.pptx
A Deep Dive on Passkeys: FIDO Paris Seminar.pptx
 
[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality Assurance[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality Assurance
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...
 
MuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotes
MuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotesMuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotes
MuleSoft Online Meetup Group - B2B Crash Course: Release SparkNotes
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
 
Zeshan Sattar- Assessing the skill requirements and industry expectations for...
Zeshan Sattar- Assessing the skill requirements and industry expectations for...Zeshan Sattar- Assessing the skill requirements and industry expectations for...
Zeshan Sattar- Assessing the skill requirements and industry expectations for...
 
Glenn Lazarus- Why Your Observability Strategy Needs Security Observability
Glenn Lazarus- Why Your Observability Strategy Needs Security ObservabilityGlenn Lazarus- Why Your Observability Strategy Needs Security Observability
Glenn Lazarus- Why Your Observability Strategy Needs Security Observability
 
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
 
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS:  6 Ways to Automate Your Data IntegrationBridging Between CAD & GIS:  6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integration
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examples
 
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxThe Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and Cons
 

Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Method 8270D Semi-Volatile Organic Compound Analysis

  • 1. a p p l i c at i o n n o t e Gas Chromatography/ Mass Spectrometry Authors Yury Kaplan Ruben Garnica PerkinElmer, Inc. Shelton, CT 06484 USA Improved Sensitivity Introduction U.S. Environmental Protection and Dynamic Range Agency (EPA) Method 8270D – Semi- Volatile Organic Compounds by Gas Using the Clarus SQ 8 Chromatography/Mass Spectrometry (GC/MS) – is a common and wide ranging GC/MS System for method employed in nearly all commercial environmental laboratories. The analysis EPA Method 8270D focuses on the detection of trace level semi-volatile organic compounds in Semi-Volatile Organic extracts from solid waste matrices, soils, air sampling media and water samples. Compound Analysis The method lists over 200 compounds however a majority of laboratories target between 60 and 90 for most analyses. The study presented here demonstrates the PerkinElmer® Clarus® SQ 8 GC/MS, not only meets the method requirements but provides users flexibility to satisfy their individual productivity demands. An extended calibration range is presented as are the advantages of the Clarifi™ detector.
  • 2. Experimental Calibration and performance standards were prepared The PerkinElmer Clarus SQ 8T GC/MS system operating in from commercially available stock standards and diluted in electron ionization mode was used to perform these experi- Class-A volumetric flasks. Calibration standards were made ments. Experimental conditions are presented in Table 1 from a 76 component 8270 mega mix, a 15 component and represent standard conditions for EPA Method 8270D. acids fortification mix and a 6 component semi-volatile Minimum inlet reactivity was achieved using a deactivated internal standard mix. The acids were fortified with the quartz liner (4-mm) without wool. As prescribed by the fortification mix to double the concentration relative to method, a pressure pulse injection scheme was employed the other components. Stock standards were diluted to and is described in Table 2. Data collection was achieved an intermediate concentration from which initial calibra- using the TurboMass™ v6 GC/MS software with environ- tion standards were prepared. The calibration levels used mental reporting. The TurboMass software comes with a in this study are presented in Table 3. The internal standard built-in UltraTune™ (Standard – DFTPP/BFB) tuning procedure, mix was added to all samples such that a resulting concen- which was used to establish the base detector setting. tration of 40 μg/mL was achieved. The tuning standard, Decafluorotriphenylphosphine (DFTPP), was prepared at 50 μg/mL. Standards, stock standards, and associated Table 1. GC and MS experimental conditions. QC/QA samples were stored in a manner consistent with Analytical Column: Elite-5MS (30 m x 0.25 mm id x 0.25 μm) the guidelines set out in the method. Injector Temperature: 300 °C Carrier Gas: Helium @ 1.0 mL/min Table 3. Calibration points employed in this study. Oven Program: Temperature Hold Time Rate “Acids” “Bases/Neutrals” 37 °C 0.5 min 18 °C/min Calibration Concentration Concentration Level (μg/mL) (μg/mL) 265 °C 0 min 6 °C/min 1 2 1 287 °C 0 min 25°C/min 2 4 2 320 °C 1.85 min End 3 8 4 GC Transfer Line Temperature: 280 °C 4 12 6 MS Ion Source 5 20 10 Temperature: 280 °C 6 30 15 MS Function Type: Full Scan 7 40 20 MS Solvent Delay: 0 – 2.30 min 8 80 40 MS Scan Range: m/z 35 – 500 9 160 80 MS Scan Time: 0.10 sec 10 200 100 MS Interscan Delay: 0.10 sec 11 300 150 Table 2. Pressure plus flow settings. Results and Discussion Time (min) Split Flow (mL/min) An initial consideration when performing any method is to determine the operating range with respect to on-column -0.50 4 concentration. The Clarus SQ 8 GC/MS allows the user to -0.25 0 control the linear range by controlling the Clarifi detector 0.75 1 voltage setting. The built-in UltraTune (Standard – DFTPP/ 1.00 50 BFB) tune procedure adjusts the mass spectrometer providing excellent spectral fidelity and sets the detector voltage such 5.00 20 that low sensitivity work can be performed. This standard 2
  • 3. setting can be used directly and is recommended for most Scan EI+ 198 7.22e8 applications. Additionally, the detector voltage can be 100 increased for added sensitivity when needed, or, as is the case with this application note, the detector voltage can be tuned to provide both sensitivity and a wide dynamic range 442 77 % 127 255 69 51 110 275 206 129 107 186 443 441 224 50 244 256 117 39 78 93 167 296 423 57 148 323 334 365 444 0 m/z 43 93 143 193 243 293 343 393 443 493 543 Figure 2. Mass spectrum of DFTPP as analyzed by the tune evaluation shown in Figure 1. suitable for the analysis at hand. In this study the Clarifi detector voltage was optimized such that both the low level samples were measurable while the high level samples were not saturating the detector. Specifically, the detector was lowered 260V to the operational 1400V. The added sensitivity of the Clarifi detector enhances a wider range of analyses so that even at low voltage settings the low concentration Figure 1. EPA 8270D tune evaluation as performed by the built-in software. Passing result shown. standards and samples are easily detected. Scan EI+ 63 TIC 100 43 32 36 23 57 58 62 54 53 25 46 60 20 27 28 71 52 29 38 2 67 72 47 12 24 56 59 % 48 50 8 13 68 69 18 41 51 10 15 65 66 31 16 55 17 30 1 0 Time 3.00 5.00 7.00 9.00 11.00 13.00 15.00 17.00 19.00 Figure 3. TIC of 40 μg/mL calibration sample. Retention Times are listed in Table 4 (Page 4). 3
  • 4. 67 Scan EI+ 252 In addition to providing a solid foundation from which to 100 perform analyses the UltraTune (Standard – DFTPP/BFB) function also produces a satisfactory tune, which meets the 68 tune evaluation requirements set out in Method 8270D. 66 Figure 1 demonstrates the passing DFTPP tune evaluation sample while Figure 2 shows the mass spectrum utilized in % the tune evaluation test. The GC conditions were optimized to provide analytical separation and Figure 3 demonstrates a mid-level (40 μg/mL) total ion chromatogram (TIC) from this analysis. All of the expected analyte separation is achieved and a listing of retention times (RT) measured using the instrument parameters 0 Time 15.00 15.20 15.40 15.60 15.80 16.00 16.20 16.40 16.60 16.80 17.00 as described in Table 1 is presented. The Elite-5MS column Figure 4. Zoom of EIC at m/z = 252 showing separation of benzo[b] provided excellent separation and Figure 4, for example, fluoranthene (66) and benzo[k]fluoranthene (67). Benzo[a]pyrene (68) demonstrates the separation of the benzo[b]fluoranthene also shown. and benzo[k]fluoranthene. Table 4. Calibration data table with retention times listed in Figure 3 (Page 3). # Name RT (min) Average RRF 8270D Criteria RRF % RSD 1 Pyridine 3.09 0.2 8.98 2 Aniline 4.44 1.2 13.16 3 Phenol 4.45 1.7 0.8 9.52 4 N-Nitrosodimethylamine 4.45 0.1 10.3 5 Bis (2-chloroethyl) ether 4.49 2.1 0.7 12.66 6 1,3-Dichlorobenzene 4.66 1.1 15.56 7 1,4-Dichlorobenzene - D4 (ISTD) 4.73 8.83 8 1,4-Dichlorobenzene 4.75 1.0 9.73 9 1,2-Dichlorobenzene 4.91 1.3 11.27 10 Benzyl Alcohol 4.94 1.0 11.86 11 Bis (2-chloroisopropyl) ether 5.05 1.7 9.34 12 2-Methylphenol 5.07 0.6 0.7 10.66 13 3-Methylphenol 5.26 0.7 11.71 14 4-Methylphenol 5.26 0.9 0.6 8.59 15 Hexachloroethane 5.28 0.5 0.3 10.62 16 Nitrobenzene 5.41 0.4 0.2 9.91 17 Isophorone 5.70 0.7 0.4 11.31 18 2-Nitrophenol 5.78 0.2 11.86 19 Bis(2-chloroethoxy)methane 5.96 0.5 0.3 15.45 20 2,4-Dichlorophenol 6.08 0.3 0.2 16.69 21 1,2,4-Trichlorobenzene 6.14 0.4 14.19 22 Naphthalene - D8 (ISTD) 6.21 7.41 23 Naphthalene 6.24 1.1 0.7 9.63 24 Hexachlorobutadiene 6.39 0.2 0.01 10.38 25 4-chloro-3-methyl-phenol 7.01 0.3 0.2 16.86 26 2-Methylnaphthalene 7.09 0.6 0.4 14.73 27 1-Methylnaphthalene 7.22 0.6 13.96 28 Hexachlorocyclopentadiene 7.30 0.2 0.05 9.14 29 2-Chloronaphthalene 7.72 1.4 0.8 12.09 4
  • 5. Table 4 continued # Name RT (min) Average RRF 8270D Criteria RRF % RSD 30 2-Nitroaniline 7.93 0.4 0.01 9.54 31 Dimethyl phthalate 8.19 1.4 0.01 10.67 32 2,6-Dinitrotoluene 8.25 0.3 0.2 13.95 33 Acenaphthylene 8.25 2.0 0.9 8.27 34 Acenaphthene - D10 (ISTD) 8.45 7.56 35 Acenaphthene 8.49 1.5 0.9 9.85 36 3-Nitroaniline 8.49 0.2 0.01 14.4 37 2,4-Dinitrophenol 8.60 0.1 0.01 8.17 38 Dibenzofuran 8.71 1.7 0.8 11.83 39 4-Nitrophenol 8.79 0.1 0.01 10.14 40 2,3,4,6-Tetrachlorophenol 8.85 0.2 12.71 41 2,3,5,6-Tetrachlorophenol 8.91 0.2 10.17 42 Diethylphthalate 9.09 1.1 0.01 9.93 43 Fluorene 9.15 1.3 0.9 11.93 44 4-Chlorodiphenylether 9.17 0.5 0.4 13.53 45 4,6-Dinitro-2-methylphenol 9.29 0.1 0.01 16.04 46 4-Nitroaniline 9.31 0.1 0.01 12.33 47 N-Nitrosodiphenylamine 9.36 0.5 0.01 11.4 48 4-Bromophenyl phenyl ether 9.80 0.2 0.1 9.57 49 Azobenzene 9.81 0.1 10.13 50 Hexachlorobenzene 9.88 0.2 0.1 10.17 51 Pentachlorophenol 10.18 0.1 0.05 8.01 52 Phenathrene - D10 (ISTD) 10.38 3.48 53 Phenanthrene 10.42 1.1 0.7 13.5 54 Anthracene 10.48 1.1 0.7 15.16 55 Carbazole 10.72 0.4 0.01 8.39 56 Di-n-butyl phthalate 11.20 0.8 0.01 11.42 57 Fluoroanthene 11.97 1.0 0.6 8.52 58 Pyrene 12.26 1.2 0.6 9.62 59 Butyl benzyl phthalate 13.16 0.4 0.01 12.35 60 Bis (2-ethylhexyl) adipate 13.26 0.5 14.38 61 Benzo[a]anthracene 13.89 1.0 0.8 11.8 62 Chrysene - D12 (ISTD) 13.92 6.34 63 Chrysene 13.96 1.0 0.7 16.57 64 Bis (2-ethylhexyl) phthalate 13.98 0.6 0.01 17.47 65 Di-n-Octyl phthalate 15.04 1.0 11.13 66 Benzo[b]fluoranthene 15.68 1.2 0.7 12.55 67 Benzo[k]fluoranthene 15.74 0.9 0.7 13.52 68 Benzo[a]pyrene 16.29 1.0 0.7 9.46 69 Perylene - D12 (ISTD) 16.42 4.24 70 Indeno(1,2,3-c,d)pyrene 18.00 1.2 0.5 8.72 71 Dibenz[a,h]anthracene 18.12 1.0 0.4 9.25 72 Benzo[g,h,i] perylene 18.42 1.0 0.5 7.98 5
  • 6. The results of the initial calibration curve for EPA Method 8270D are presented in Table 4 and include retention time, average relative response factor (Avg. RRF), method criteria, and percent relative standard deviation (% RSD). Satisfactory results are obtained for all for the compounds listed. Conclusion In this application note the Clarus SQ 8 GC/MS system is shown to provide wide concentration range applicability to U.S. EPA Method 8270D, allowing the analysts flexibility in performing experimentation to meet their needs. Satisfactory analytical results were demonstrated over a concentration range from 1.0 – 150 μg/mL for a majority of analyte compounds using the full scan approach providing library searchable spectra at all concentration levels. A number of technological advances make the Clarus SQ 8 GC/MS the ideal systems for laboratories wishing to perform high throughput and sensitivity analyses with an ease of operation that is currently unmatched. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010016_01