Atomic Absorption Spectrophotometer is for determining the concentration of a particular metal(element) in a sample. Atomic Absorption Spectrophotometer can be used to analyze the concentration of over 62 different metals. For More Information Please Logon http://goo.gl/1dTEGl
2. R
ATOMIC ABSORPTION
SPECTROPHOTOMETER
Atomic Absorption Spectrophotometer is for determining the concentration of a particular metal(element) in a
sample. Atomic Absorption Spectrophotometer can be used to analyze the concentration of over 62 different metals.
The technique makes use of absorption spectrometry to assess the concentration of an analyte in a sample. It relies
therefore heavily on Beer-Lambert law. In short,
the electrons of the atoms in the atomizer can be
promoted to higher Orbitical's for an instant by
absorbing a set quantity of energy (i.e. light of a
given wavelength). This amount of energy (or
wavelength) is specific to a particular electron
change in a particular element, and in general, each wavelength
corresponds to only one element. This gives the technique its elemental
selectivity. As the quantity of energy (the power) put into the flame is known, and the quantity
remaining at the other side (at the detector) can be measured, it is possible, from Beer-Lambert law, to calculate how
many of these transitions took place, and thus get a signal that is proportional to the concentration of the element being
measured.
TECHNIQUES
In order to analyze a sample for its atomic constituents, it has to be atomized. The sample should then be illuminated by
light. The light transmitted is finally measured by a detector. In order to reduce the effect of emission from the
atomizer(e.g. the black body radiation) or the environment, a spectrometer is normally used between the atomizer and
TECHNOLOGIES PVT. LTD.
3. R
the detector. The technique typically makes use of a flame to atomize the sample,[3] but other atomizers such as a
graphite furnace[4] or plasmas, primarily inductively coupled plasmas, are also used.[5]When a flame is used it is
laterally long (usually 10 cm) and not deep. The height of the flame above the burner head can be controlled by
adjusting the flow of the fuel mixture. A beam of light passes through this flame at its longest axis (the lateral axis) and
hits a detector.
ANALYSIS OF LIQUIDS
A liquid sample is normally turned into an atomic gas in three steps:
Ÿ Desolvation - the liquid solvent is evaporated, and the dry sample remains
Ÿ Vaporization - the solid sample vaporises to a gas
Ÿ Atomization - the compounds making up the sample
FEATURES
Ÿ Built-in computer data processing and LCD:adopting highly integrated micro digital circuit, which is stable and
reliable with the functions of integral holding, peak height and area, auto zero adjusting, deuterium lamp and
background, multi-linear and nonlinear curves fitting, various parameters and working curves displayed in screen
and report printing, etc. It is equipped with interface for externally linking PC.
Ÿ Good basic line stability:Double-beam system of optimized design can automatically compensate the light source
drift and wavelength drift caused by the variation of temperature (with the function of the eliminating the affection of
wavelength drift on the base line stability) and electronic circuit drift so as to reach a good basic line stability. The
cathode lamp needs not be pre-heated for long time and sample can be analyzed immediately.
TECHNOLOGIES PVT. LTD.
4. R
the detector. The technique typically makes use of a flame to atomize the sample,[3] but other atomizers such as a
graphite furnace[4] or plasmas, primarily inductively coupled plasmas, are also used.[5]When a flame is used it is
laterally long (usually 10 cm) and not deep. The height of the flame above the burner head can be controlled by
adjusting the flow of the fuel mixture. A beam of light passes through this flame at its longest axis (the lateral axis) and
hits a detector.
ANALYSIS OF LIQUIDS
A liquid sample is normally turned into an atomic gas in three steps:
Ÿ Desolvation - the liquid solvent is evaporated, and the dry sample remains
Ÿ Vaporization - the solid sample vaporises to a gas
Ÿ Atomization - the compounds making up the sample
FEATURES
Ÿ Built-in computer data processing and LCD:adopting highly integrated micro digital circuit, which is stable and
reliable with the functions of integral holding, peak height and area, auto zero adjusting, deuterium lamp and
background, multi-linear and nonlinear curves fitting, various parameters and working curves displayed in screen
and report printing, etc. It is equipped with interface for externally linking PC.
Ÿ Good basic line stability:Double-beam system of optimized design can automatically compensate the light source
drift and wavelength drift caused by the variation of temperature (with the function of the eliminating the affection of
wavelength drift on the base line stability) and electronic circuit drift so as to reach a good basic line stability. The
cathode lamp needs not be pre-heated for long time and sample can be analyzed immediately.
TECHNOLOGIES PVT. LTD.
5. R
It is the preferable instrument chosen by users to conduct analysis of multiple kinds of elements and fast analysis of
samples.
Ÿ High precision of measurement:Gas path system is equipped with precision pressure stabilizing and current
stabilizing devices to reach stable flame and low noise. Specially designed fine light beam passes through the
flame to ensure a high precision analytical test and low characteristic concentration.
Ÿ Optical path of high energy: A total reflection system is adopted to eliminate color difference in full range. By means
of chemical conversion, a round light spot of the light source becomes a long light spot, which enters into the slit.
Therefore the light flux of double beam is enhanced.
Ÿ Long-life and anti-corrosive atomization system: The burner is made of new type titanium alloys, anti-corrosive and
fast thermal equilibrium. It meets the requirement of measurement sensitivity without water-cooling.
Ÿ Multi-functional analysis mode: for methods of flame absorption, flame emission, graphite furnace atomic
absorption (AS-3020 automatic sample feeder can be installed) and hydride generation.
Ÿ Safe and reliable gas path system: Special devices of quick gas conversion and safety protection can be used to
analyze air-acetylene flame as well as nitrous oxide-acetylene flame and extend the analytic elements to reach
more than 60.
Ÿ Complete set of accessories: to be supplied with the instrument and ready for use after they are purchased.
TECHNIQUES
Ÿ In order to analyze a sample for its atomic constituents, it has to be atomized. The sample should then be
illuminated by light. The light transmitted is finally measured by a detector. In order to reduce the effect of emission
from the atomizer (e.g. the black body radiation) or the environment, a spectrometer is normally used between the
atomizer and the detector.
TECHNOLOGIES PVT. LTD.
6. R
Characteristic Concentration
Characteristic concentration of copper = 0.04µg/ml/1%
Detection Limit
Detection limit of copper ≤0.008μg/ml
Background Calibration Ability
Greater than 30 times
Light-Source System
Hollow cathode lamp power supply: with adjustable current Deuterium Lamp
power supply: with fixed current
Optical system
Double beam total refection system, C-T type monochromator Flashing
wavelength: 250nm
Flame atomization system
Inter-changeable titanium alloy burner of 10cm and 5cm slits. Gas path is equipped
with pressure indication, high precision voltage stabilizing flow adjustment, fast
conversion of air and nitrous oxide, gas supply cut off, power cut off, prevention of
flashback and logic interlock protection devices.
Signal processing
Build-in microprocessor and LCD. Signal amplifying modes: absorptance,
concentration, emission and emission concentration; Standard curves: linear
regression, curve fitting and multi-point standard calibration. Average value,
standard deviation, relative standard deviation and correlation coefficient can be
worked out. Standard curves, atomic absorption peak diagram, spectrum outline
sketch, data as well as instrument parameters list and analytical reports can be
displayed and printed out.
Standby Interface
TECHNOLOGIES PVT. LTD.
RS 232 or USB
7. R
Instrument power supply
220V, 3A, 50Hz
Instrument Dimension
1000mmx530mmx425mm
Instrument Weight
130kg
Accessories Supplied for (Atomic Absorption Spectrophotometer)
Main accessories supplied
Several decades of accessories including air compressor, printer, one piece of
with the instrument
(Zn/Cu) and K. Gas external pipe, acetylene cylinder pressure-reducing valve and
water knockout gas filter.
As3020 graphite furnace
Used to improve automation and measurement precision
automatic sample feeder
HGA graphite furnace system
Used to conduct trace and ultra-trace analysis with an absolute sensitivity of 10-1010-12 gram.
5cm long-life titanium burner
The necessary accessory for nitrous oxide-acetylene flame analysis.
Hydride generator (Totally
Making the characteristic concentration of As, Se, Sb, Bi, Pb, Sn, Te, Ge, In, Tl, Cd
automatic)
and Zn lower than ppb level with cold atomic mercury measurement method.
Nitrous oxide generator
The necessary accessory for nitrous oxide-acetylene flame analysis.
(Totally automatic)
Hollow cathode lamp
Graphite tube
All specifications are available.
All specifications are available.
TECHNOLOGIES PVT. LTD.
8. R
Instrument power supply
220V, 3A, 50Hz
Instrument Dimension
1000mmx530mmx425mm
Instrument Weight
130kg
Accessories Supplied for (Atomic Absorption Spectrophotometer)
Main accessories supplied
Several decades of accessories including air compressor, printer, one piece of
with the instrument
(Zn/Cu) and K. Gas external pipe, acetylene cylinder pressure-reducing valve and
water knockout gas filter.
As3020 graphite furnace
Used to improve automation and measurement precision
automatic sample feeder
HGA graphite furnace system
Used to conduct trace and ultra-trace analysis with an absolute sensitivity of 10-1010-12 gram.
5cm long-life titanium burner
The necessary accessory for nitrous oxide-acetylene flame analysis.
Hydride generator (Totally
Making the characteristic concentration of As, Se, Sb, Bi, Pb, Sn, Te, Ge, In, Tl, Cd
automatic)
and Zn lower than ppb level with cold atomic mercury measurement method.
Nitrous oxide generator
The necessary accessory for nitrous oxide-acetylene flame analysis.
(Totally automatic)
Hollow cathode lamp
Graphite tube
All specifications are available.
All specifications are available.